Speech Generating Device Recall: Tobii Dynavox Navio Maxi 1000799
Announced 2026-08-26 · FDA
An FDA recall covers the Tobii Dynavox Navio Maxi Speech Generating Device (model 1000799). The recall concerns damaged battery mounting tabs that may allow the battery to move within the enclosure, potentially causing an internal short circuit with localized heating and gas venting. Affected serial numbers are TDNL1-240904000016 through TDNL1-260617070001; corrected/future production will have new serialization from TDNL1-260617070001 onward (production batch 20260615, improved design).
| Brand | Tobii Dynavox Llc |
|---|---|
| Product | TD Navio Maxi Speech Generating Device Model/Catalog Number: 1000799 |
| Reason | Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure. |
| Severity | moderate |
| Distribution | US Nationwide distribution. |
| Agency | FDA |
| Recall / campaign no. | UDI-DI 073400747602070 Serial Numbers TDNL1-240904000016 through TDNL1-260617070001. Corrected/future production is differentiated by new serialization from TDNL1-260617070001 onward (production batch 20260615, improved design). |
WHAT YOU SHOULD DO
- If your device's serial number falls within the affected range, stop using it until you receive instructions from Tobii Dynavox.
- Contact Tobii Dynavox customer support for recall instructions, inspection, repair, or replacement options.
- Follow the recall guidance from FDA and Tobii Dynavox for return or repair and do not attempt DIY repairs.
- Consider using an alternate communication method while awaiting resolution.