Battery Kit Recall: Philips Respironics SimplyGo Mini

Announced 2026-04-08 · FDA

Certain serial numbers of the SimplyGo Mini standard battery kit (Product code: 1116816) contain a defective chip that can cause the battery to fail to charge or lose voltage during use. If this occurs while the device is running on battery, the SimplyGo Mini could shut down and stop delivering oxygen. The recall covers 56 units distributed in the US (GA, ID, KY, NJ, NY) and Canada.

BrandPhilips Respironics, Inc.
ProductSimplyGo Mini, Standard Battery Kit; Product code: 1116816; The affected battery kits are sold as accessories for the SimplyGo Mini: SimplyGo Mini Portable Oxygen Concentrator Product code: 1113600
ReasonCertain serial numbers of the SimplyGo Mini standard battery kit contain a defective chip which can lead the battery to either fail to charge or lose voltage output when in use in the device. If the fault occurs while the device is running on battery power, the SimplyGo Mini will shut down and stop providing oxygen to the user. This may happen at a temperature within the specified operating temperature range of the SimplyGo Mini.
Severitymoderate
Units affected113717
DistributionUS Nationwide distribution in the states of GA, ID, KY, NJ, NY and the country of Canada.
AgencyFDA
Recall / campaign no.Product Code: 1116816; UDI-DI: 00606959032385; Serial Numbers: 2438V115830, 2438V115960, 2438V116033, 2438V115899, 2438V115832, 2438V115964, 2438V116034, 2438V115837, 2438V115965, 2438V116040, 2438V115840, 2438V115967, 2438V116044, 2438V115842, 2438V115968, 2438V116045, 2438V115844, 2438V115970, 2438V116046, 2438V115878, 2438V115972, 2438V116047, 2438V115890, 2438V115973, 2438V116048, 2438V115894, 2438V115974, 2438V116049, 2438V115895, 2438V115980, 2438V116050, 2438V115896, 2438V115981, 2438V116051, 2438V115897, 2438V115983, 2438V116052, 2438V115919, 2438V115984, 2438V116226, 2438V115937, 2438V115991, 2438V116229, 2438V115939, 2438V115992, 2438V116233, 2438V115942, 2438V115993, 2438V116249, 2438V115948, 2438V115994, 2438V116253, 2438V115957, 2438V115995, 2438V116258, 2438V117762;

WHAT YOU SHOULD DO

  1. Stop using the affected battery kit and remove it from the SimplyGo Mini.
  2. Contact Philips Respironics to arrange replacement or refund; provide your product code (1116816) and the affected serial numbers to verify eligibility.
  3. If your device shuts down, switch to an alternative power source and contact your healthcare provider for guidance.
  4. Keep documentation of the recall and your communications with Philips Respironics for reference.

This is an automatically maintained record of an official recall notice. Details can change; always follow the official notice linked above. See all records in the Recall Database.