Blood collection tubes Recall: SAFE-T-FILL Capillary
FDA recall for SAFE-T-FILL Micro Capillary Blood Collection tubes manufactured between July 31, 2023 and February 28, 2025. They may cause false positive lead results when used with Magellan Diagnostics LeadCare Testing Systems, potentially delaying results and prompting unnecessary testing. About 4,272,858 units were distributed worldwide, including across the US.
| Brand | ASP Global, LLC. dba Anatomy Supply Partners, LLC. |
|---|---|
| Product | SAFE-T-FILL Micro Capillary Blood Collection, 125 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 6011; SAFE-T-FILL Micro Capillary Blood Collection, 125 ¿L Prepared with Dipotassium EDTA Self-sealing Cap, Purple, Model: 07 6013; SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7051; SAFE-T-FILL Micro Capillary Blood Collection, 150 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7052; SAFE-T-FILL Micro Capillary Blood Collection, 300 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7053; SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7056; SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L Prepared with Dipotassium EDTA Flat Bottom Microtube, Purple, Model: 07 7058 SAFE-T-FILL¿ End-to-End Capillary Tubes, 30 ¿L; Plastic, Prepared with EDTA Dipotassium Model: 06 0910 |
| Reason | All RAM SAFE-T-FILL Micro Capillary Blood Collection tubes manufactured between July 31, 2023 to February 28, 2025 cause false positive results when used with Magellan Diagnostics LeadCare Testing Systems. False positive lead results may cause delayed results and additional unnecessary testing. |
| Severity | moderate |
| Units affected | 620502 |
| Distribution | Worldwide - US Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, PA, SC, SD, TN, TX, UT, VA, WA, WI and the country of CA. |
| Agency | FDA |
WHAT YOU SHOULD DO
- If you had a lead test performed with these tubes, ask your healthcare provider to review the results for possible false positives.
- If you are a facility using any of the listed lots or models, stop using them and contact ASP Global for guidance on replacement or recall procedures; quarantine affected stock per your protocols.
- Check your records for the listed models and lots (07 6011, 07 6013, 07 7051, 07 7052, 07 7053, 07 7056, 07 7058, and 06 0910) to determine if you are affected.
- Monitor updates from the FDA and the manufacturer for further instructions and testing recommendations.
How to identify the recalled item
| Lot / date codes | Model/UDI-DI/Lots: 07 6011/10643351000014/23H4110, 23I4017, 23I4051, 23J4031, 23K4058, 23K4059, 23L4082, 24A4103, 24B4186, 24B4187, 24D4015, 24E4138, 24E4139, 24H4121, 24I4317, 24I4328, 24J4041, 24J4137, 24J4138, 24J4233, 24J4316, 24J4348; 07 6013/10643351000021/23H4111, 23I4018, 24B4188, 24H4125, 25B4099; 07 7051/10643351000038/23H4157, 23I4020, 23I4374, 23J4033, 23K4060, 23L4083, 24A4104, 24B4190, 24D4016, 24E4140, 24F4053, 24I4321, 24J4043, 24J4139, 24J4237, 24L4001, 25B4107; 07 7052/10643351000045/23H4113, 23I4021, 23J4034, 24A4105, 24D4017, 24F4052, 24H4126, 24I4323, 24J4042, 24J4140, 24J4338, 24J4355, 25B4108; 07 7053/10643351000052/23I4022, 23J4035, 23L4084, 24D4018, 24E4132, 24F4192, 25B4106; 07 7056/10643351000069/24A4106, 24B4191, 24B4192, 24E4133, 24I4315, 24J4136, 24J4347; 07 7058/10643351000076/23G4430, 23G4168. Manufactured between July 31, 2023 and February 28, 2025. |
|---|---|
| Lot / date codes | Model Number: 06 0402 GTIN (Inner Pack): 30643351000346 UDI-DI: 10643351000342 All Lots produced from 01AUG2020 to 28FEB2025: 22A4541, 22G4485, 22K4018, 23A4160, 23C4340, 23F4168, 23F4169, 23H4107, 23I4055, 23J4039, 24E4129, 24F4189 |
These are the identifiers FDA published for this recall. Most recall notices list no UPC or model number at all — if the code on your item is not shown here, that does not mean the item is excluded. Check the official notice, or contact the company.
This is an automatically maintained record of an official recall notice. Details can change; always follow the official notice linked above. See all records in the Recall Database.